Consider the following statements regarding sources of error in a Q-meter:
1. If a coil with resistance R is connected in direct measurement mode and if the residual resistance of Q-meter is 0.1 R, then the measured Q of the coil would be 1.1 times the actual Q.
2. If the inductance to be measured is less than 0.1 mH, the error due to presence of residual inductance cannot be neglected.
3. The presence of distributed capacitance in a coil modifies the effective Q of the coil.
Which of the above statements are correct?
A. 1, 2 and 3
B. 1 and 2 only
C. 2 and 3 only
D. 1 and 3 only
Answer: Option C
Related Questions on Measurement and Instrumentation
The spring material used in a spring control device should have the following property.
A. Should be nonmagnetic
B. Most be of low temperature coefficient
C. Should have low specific resistance
D. All of the above
Why is a MISC meter not recommended for DC measurement?
A. The meter is calibrated for AC and it’s error for DC would be high
B. The meter does not respond to DC signals
C. The error is high due to hysteresis effect
D. The error is high due to eddy current effect
A moving iron ammeter may be compensated for frequency errors by
A. A shunt resistance
B. A series inductance
C. Shunt capacitance
D. Series resistance