Orsat apparatus is meant for
Answer & Solution
Correct Answer:
Option
D
The Orsat apparatus is used for the volumetric analysis of flue gases produced during combustion. It determines the percentage composition of different gases present in the flue gas by using selective chemical absorption.
The three major gases commonly analyzed using the Orsat apparatus are:
1. Carbon Dioxide (CO₂): Absorbed by Potassium Hydroxide (KOH).
2. Oxygen (O₂): Absorbed by alkaline pyrogallol.
3. Carbon Monoxide (CO): Absorbed by ammoniacal cuprous chloride.
What about Nitrogen (N₂)?
Nitrogen is not directly determined by absorption in the Orsat apparatus. Its percentage is generally calculated by difference after measuring the other gas constituents.
Why are the given options incorrect?
Option A: Gravimetric analysis of flue gas – Incorrect.
Gravimetric analysis is based on measuring the mass of a substance. The Orsat apparatus works on the principle of volumetric gas analysis, not gravimetric analysis.
Option B: Finding out combustion efficiency – Incorrect.
The Orsat apparatus does not directly measure combustion efficiency. It determines the volumetric composition of flue gases, and the obtained results can subsequently be used to assess combustion performance.
Option C: Direct determination of nitrogen in flue gas by absorbing it in ammoniacal cuprous chloride – Incorrect.
Ammoniacal cuprous chloride selectively absorbs carbon monoxide (CO), not nitrogen (N₂). Nitrogen is generally determined by difference.
Therefore, none of the given options correctly describes the primary function of the Orsat apparatus. Hence, Option D (None of these) is the correct answer.
The three major gases commonly analyzed using the Orsat apparatus are:
1. Carbon Dioxide (CO₂): Absorbed by Potassium Hydroxide (KOH).
2. Oxygen (O₂): Absorbed by alkaline pyrogallol.
3. Carbon Monoxide (CO): Absorbed by ammoniacal cuprous chloride.
What about Nitrogen (N₂)?
Nitrogen is not directly determined by absorption in the Orsat apparatus. Its percentage is generally calculated by difference after measuring the other gas constituents.
Why are the given options incorrect?
Option A: Gravimetric analysis of flue gas – Incorrect.
Gravimetric analysis is based on measuring the mass of a substance. The Orsat apparatus works on the principle of volumetric gas analysis, not gravimetric analysis.
Option B: Finding out combustion efficiency – Incorrect.
The Orsat apparatus does not directly measure combustion efficiency. It determines the volumetric composition of flue gases, and the obtained results can subsequently be used to assess combustion performance.
Option C: Direct determination of nitrogen in flue gas by absorbing it in ammoniacal cuprous chloride – Incorrect.
Ammoniacal cuprous chloride selectively absorbs carbon monoxide (CO), not nitrogen (N₂). Nitrogen is generally determined by difference.
Therefore, none of the given options correctly describes the primary function of the Orsat apparatus. Hence, Option D (None of these) is the correct answer.